6 Watlow Series V4
Input Wiring
Figure 6a – Thermocouple
Extension wire for thermocouples must be of the same alloy as the thermo-
couple itself to limit errors.
Figure 6b – RTD (2- or 3-Wire) 100Ω Platinum
There could be a +2°F input error for every 1Ω of lead length resistance
when using a 2-wire RTD. That resistance, when added to the RTD element
resistance, will result in erroneous input to the instrument. To overcome
this problem, use a 3-wire RTD sensor, which compensates for lead length
resistance. When extension wire is used for a 3-wire RTD, all wires must
have the same electrical resistance (i.e. same gauge, same length, multiple-
stranded or solid, same metal.
Figure 6c – Process,4-20mA Figure 6d – Process,0-5VÎ (dc)
Input impedance: 5Ω Input impedance: 10kΩ
Komentarze do niniejszej Instrukcji